10.5.3 PR‑23‑GC parts lis
t
Sandvik 76.1 reducing ferrule
2
1
PR-9282
Sandvik clamp FCLC-76.1
Sandvik O-ring FCLG-T-76.1 Teflon®
Sandvik O-ring FCLG-V-76.1 Viton
4
1
PR-9284
PR-23/33-GC head
PR-10002 O-ring seal 82x3
PR-10047-SC PR-23 endplate with label and screws
PR-10036 H73 compact sensor CORE module
Figure 61 PR‑23‑
GC parts list
10.5.4 PR‑23‑GC mounting specifications
The compact refractometer is mounted either in a pipe elbow by an L coupling connection
or in a straight pipe using a wafer flow cell or a pipe
flow cell. Both flow cell mounting
designs create optimum flow velocity on the measurement surface providing a good self-
cleaning
eect. The wafer flow cell has an optional automatic wash system capability.
In a pipe of at least 2.5 in diameter the sensor is mounted in a pipe elbow with an L coupling.
In a 2 in pipe the sensor is mounted in a pipe elbow using reducing ferrule PR-9283. In 0.5 in,
1 in and 1.5 in pipes a
flow cell is installed in a straight pipe. The flangeless wafer flow cell is a
compact alternative to traditional flow cells. The wafer refers to a flow cell body that is
installed between DIN, ANSI or JIS piping flanges with bolts and nuts. The wafer flow cell is a
one
‑piece body construction with no welds. A pipe flow cell is also available for 0.5 in and
1 in pipes.
PR-23 Series User Guide
IM-EN-PR23-E
126