To activate (or deactivate) a temperature limit, choose 7 TEMP LIMIT ACTIVATION and
then the appropriate command in the menu.
To set a low temperature limit, choose 8 TEMP LIMIT VALUE °C and enter the temperature
(in °C) where the limit should be.
The empty pipe check prevents washing if message is NO SAMPLE, meaning there is no
process liquid in the pipe. To deactivate (or active) the empty pipe check, choose 9 EMPTY
PIPE CHECK and then the appropriate menu command.
To change the wash n
D
limit, select 0 MORE … and then 1 WASH nD LIMIT to set the n
D
value to be used with the wash check functionality.
To set wash tolerance time, select 0 MORE … and then 2 WASH TOLERANCE TIME to set
the time during which a wash must be accepted. If no wash is accepted during wash
tolerance time, the message becomes PRISM WASH FAILURE. The wash tolerance counter is
reset by a successful wash.
More information
Recommended wash pressures and times (page 37)
Chapter 7 – Configuration and calibration
83