EasyManuals Logo

YOKOGAWA ADMAG AXR User Manual

YOKOGAWA ADMAG AXR
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
<6. PARAMETER DESCRIPTION>
6-34
IM 01E30D01-01EN
[J22: Dead Time] Setting of dead time
This parameter sets the time for application of the
rate limit, and if a value of 0 is set, the rate limit
function will be terminated.
NOTE
Determining rate limit value and dead time
F0611.ai
T0
T0
2%
2%
Rate limit value:
Determines the level for
output fluctuation cutoff. For
example, if this is set to 2%,
noise above 2% will be
eliminated as shown in the
diagram.
Dead time (T
0
):
This is to be determined using
the output fluctuation width. If
noise exceeds the dead time
as shown in the diagram
below, the dead time should
be made longer.
Signal processing method:
Axedupperandlowerlimitvalueissetupwith
respect to the primary delay response value for
theowratevalueobtainedduringtheprevious
sampling,andifthecurrentlysampledowrate
is outside these limits, then the corresponding
limitisadoptedasthecurrentowratevalue.
In addition, if signals which breach the limits in
the same direction occur over multiple samples
(i.e., within the dead time), it is concluded that
thecorrespondingsignalisaowratesignal.
Example 1: Step input
F0612.ai
10
%
1%
Step signal
Flow rate value
after rate limit
processing
63.2
%
100 225
Dead time: 3 s
Input: 0 to 10%
Damping time constant: 5 s
Dead time: 3 s
Rate limit value: 1%
Number of signal samples
Flow rate value after
damping
(a)
(b)
(c)
(d)
(1) In comparison with the previous value at (a), it
is determined that the signal is in excess of the
rate limit value and the response becomes 1%.
However, the actual output applies damping,
and therefore the output turns out to be as
indicated by the solid line.
(2)Subsequentowvalueswithinthedeadtime
zone correspond to signals of post-damping
owvalue+ratelimitvalue(1%).
(3) Since input signals do not return to within
the rate limit value during the dead time, it is
determinedat(c)thatthissignalisaowrate
signal.
(4) The output signal becomes a damped curve
and compliance with the step signal begins.
Fivesecondsafterdeterminationofaowrate
signalintheabovegure,alevelof63.2%is
reached.
Example 2: Slurry noise
F0613.ai
+1%
-1%
Dead time: 1 s
Flow rate value after damping
Flow rate value
after rate limit
processing
Slurry noise
Input: 0 to 10%
Damping time constant: 1 s
Dead time: 1 s
Rate limit value: 1%
Time
In the figure on the left, it
is determined that the
slurries noise signal is not
a flow rate signal.
[J23: Pulsing Flow]Selectionofpulsingow
support
Inasituationwherepulsatingowcauseserrorin
theaverageowvalue,duetotheapplicationofa
plunger pump, this parameter provides functionality
whereby calculation is controlled and variations in
owratearefollowed.
Setting Function
No Normal
Yes Support for pulsingow
[J25:T/PDampSelect] Setting of damping
operation
Thisparameterisusedtoselectthattheowrate
value obtained through damping calculation for
totalization and pulse output or the instantaneous
owratevalue(nodamping)fortotalizationand
pulse output.
Setting Function
Damp Damping
No Damp No damping

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ADMAG AXR and is the answer not in the manual?

YOKOGAWA ADMAG AXR Specifications

General IconGeneral
BrandYOKOGAWA
ModelADMAG AXR
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals