ZEISS INTRODUCTION Axio Imager 2
Notes on device safety
10 430000-7544-001 01/2016
For stands equipped with motorized focusing drives, there is a risk of fingers being crushed
in the working area when the microscope stage is lowered.
Do not reach into the area between the microscope stage / stage carrier / condenser
carrier and the stand base while the stage is being lowered.
The automatic movement can be stopped by pressing the STOP button on the TFT
display. (See Fig. 2).
When a scanning stage (approx. weight: 4 kg) is used, the high-
performance focus must be
activated to protect the focusing drive (see page 53
). Sample weights of up to 5 kg can be
placed on the stage if the high-performance focus is activated.
It is essential to read the safety data sheet provided with Immersol 518 F
immersion oil.
Immersol 518 F
immersion oil is a skin irritant. Avoid contact with skin, eyes and clothing.
In the event of contact with the skin, rinse with plenty of water and soap.
In the event of contact with the eyes, rinse the eyes
immediately with plenty of water for at
least five minutes. If irritation persists, consult a physician.
Correct disposal of Immersol 518 F
immersion oil. Ensure that the immersion oil does not enter
surface water or the sewage system.
Do not
replace detachable power cables with power cables that do not meet the specifications.
Only the specified power cables should be used.
The Axio Imager.Z2 or Z2m can only be disconnected from the mains supply using the on/off
switch on the external VP232-2 power unit.