Axio Imager 2 OPERATION ZEISS
Axio Imager 2 operation and function controls (motorized version)
01/2016 430000-7544-001 89
Light intensity button (Fig. 76/34)
− Reduces the light intensity of halogen illuminator HAL; function corresponds to Fig. 76/14.
Light intensity button (Fig. 76/35)
− Increases the light intensity of halogen illuminator HAL; function corresponds to Fig. 76/14.
Toggle switch for transmitted / reflected-light halogen illuminator (Fig. 76/36)
− Alternately switches the halogen illuminator for transmitted light or reflected light on / off.
− Toggle switch up: Reflected-light halogen illuminator ON (transmitted-light OFF)
− Toggle switch down: Transmitted-light halogen illuminator ON (reflected-light OFF)
Button LM-Set (Fig. 76/37)
− Saves the values of the current light intensity and the optical path nosepiece position (short beep).
Lever for reflected-light diffusion disk (Fig. 76/38)
− Lever in top position: Diffusion disk in light path
− Lever in bottom position: Diffusion disk not in light path
4.2 Axio Imager 2 operation and function controls (motorized version)
In this section, the operation and function controls of the motorized version of the Axio Imager are
described that are different from the manual version. Controls not described here correspond to the
manual version.
As many components are offered as options, the structure of your microscope may differ from the
following description. It is also possible, for instance, to use manually controlled components (such as
reflector turret, condensers, filter wheels) on the motorized stand. In this case, however, the functionality
of the microscope, particularly its operation via TFT display (touchscreen) will be restricted.
Precise
configuration is essential for the correct functioning of the motorized instrument,
especially the light source(s) used, objective magnifications as well as reflector modules and
filters.
Most of the settings can be made directly on the TFT display (see Section 4.8.6 on page 126).
More specific and less frequently used configurations, such as a special assembly of the
condenser or your own filter sets in the double filter wheel, can be made using the
configuration software MTB 2011 config.exe.