Axio Imager 2 FIRST-TIME SET-UP ZEISS
Mechanical stage 75x50 mot. CAN
01/2016 430000-7544-001 77
3.37 Mechanical stage 75x50 mot. CAN
To position the 75x50 mot. CAN stage,
the coaxial drive,
the trackball or
the joystick
can be used.
3.37.1 Assembling and connecting 75x50 mot. CAN mechanical stage with coaxial drive
On delivery, the mechanical stage has a transport lock. This should be removed before attaching
the stage, as otherwise the mechanical stage could be damaged.
Unpack the mechanical stage and remove the transport lock (Fig. 70/2).
Slide the stage plate (Fig. 70/1) forwards and fasten coaxial drive (Fig. 70/4) to the underside of the
stage using two screws (Fig. 70/5).
Place mechanical stage (Fig. 70/1) onto the stage carrier so that the holes in the underside of the
mechanical stage coincide with the through holes of the stage carrier.
Using the angled Allen key (AF 3), screw four screws (Fig. 70/6) from the bottom into the underside of
the stage, with the shorter screws being inserted in the front holes.
1 Mechanical stage 75x50 mot. CAN
2 Transport lock (screw)
3 Angled CAN bus connector of the connection cable (100-0600-144)
4 Electronic coaxial drive CAN
5 Fastening screws on coaxial drive
6 Fastening screws on stage carrier
Fig. 70 Assembling Mechanical Stage 75x50 mot. CAN