February 2020TD 266 OPERATING MANUAL ZPulse® DCS 4420/4830/4520/4930 Page 77
The Measurement Area is defined as the
sizeCellLengthPulse
+
The Cell Start is defined as:
2
LengthPulse
ance
dist
Start +
The Cell End is defined as:
sizeCellstartCell
+
The Cell Centre is defined as:
2
SizeCellLengthPulse
nceadistStart
+
+
Figure 5-6: Illustration of Pulse, Cell and the Measurement area.
Enable ZPulse Operation. Activate/deactivate ZPulse operation; Set yes to activate. Default value is yes. ZPulse
operation gives two separate frequencies in each ping and therefor higher quality data without increasing power
consumption