118 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Setting system modules
Menu path
MENU Configure System Set System Modules
Function
If system modules are added to an analyzer system or replaced (changed) or
removed, this also needs to be configured in the software.
Definition
System modules are the analyzer modules, the I/O modules and the external
I/O devices.
Analyzer modules and external I/O devices
The analyzer modules and external I/O devices are connected via the system
bus with the system controller. For them to be identified by the analyzer
system, they need to be identified using their serial number (see below).
I/O modules
The I/O modules are placed on the system controller and connected to it
directly. They do not have a serial number.
An I/O module is automatically identified by the analyzer system if it is new
or has already been added as a replacement for an already existing I/O
module.
Serial number
The module’s 14-digit serial number is stated in the Analyzer Data Sheet and
on the sticker on the module; the sticker is usually on the CPU card. The seri-
al number contains the following information (example):
01400000012301 (analyzer module)
13000000001600 (u-remote module, bus address 16 or H=1 L=1)
The first 3 digits refer to the module type:
014
Analyzer module
130
u-remote module