78 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Automatic Adjustment Check (AAC): Configuring the time sequence
Function
Optionally, the cycle that automatically monitors the adjustment (Automatic
Adjustment Check – AAC) of the FTIR spectrometer under QAL3 monitoring
can be configured. A validation filter wheel equipped with the AAC measure-
ment media must be installed in the FTIR spectrometer and the "Automatic
Adjustment Check" function must be enabled in the configuration file.
In the menu described here, only the time sequence of an AAC is controlled.
The QAL3 monitoring can only be configured and evaluated via the web page
(see page 80).
Menu path
Menu Configure Adjustment Data
Automatic Adj. Check
Display
The recommended settings are stated in the following table.
Procedure
Parameter Explanation
Enable
The AAC is only completed when it is activated.