86 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Configuring automatic adjustment of the FID
Function
For the FID, an automatic adjustment with test gas at the zero and end point
can be configured.
Menu path
Menu → Configure → Adjustment Data →
Automatic Adj.
Display
The recommended settings are stated in the following table.
Procedure
Parameter Explanation
Enable
Automatic adjustment is completed only when it is activated.
Cycle time
The cycle time shows the time intervals over which automatic adjustment is
completed. Recommended setting: 21 days
Date/time of next
adjustment
The analyzer system completes the next automatic adjustment at the time
specified here. From this moment in time, the cycle period starts to run.