170 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Replacing the battery on the system controller
Wearing parts
Wearing parts Type Replacements
Battery on the
system controller
Varta CR3032 (type no.
6032) or Renata CR2032
Approx. every 4 years (as part
of the maintenance work on
the analyzer system)
CAUTION! Located on the system controller are electronic components, which are sus-
ceptible to damage by electrostatic discharges.
Measures for preventing electrostatic discharges must be taken during bat-
tery replacement.
Before the batteries are replaced, make sure that the power supply to the
analyzer system is switched off.
Replacing the battery on the system controller
1 Slacken 10 screws and remove the cover of the system controller hous-
ing.
2 Remove used battery 1 from the holder and insert new battery into the
holder.
3 Mount cover of the system controller housing and fasten using the
screws.
4 Dispose of used battery in accordance with the locally applicable regula-
tions.