120 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Add system module
NOTICE
As long as no system module is configured in the analyzer system at all or an
added system module is not yet configured, the NEW softkey appears in the
display in measuring mode. By pressing this softkey, the user goes directly
to the Set system modules menu.
The approach when adding an analyzer module or an external I/O device
differs from that when adding an I/O module (see following instructions).
Add a new analyzer module or a new I/O device
1 Select Set system modules menu item.
The list of the system modules available in the system is displayed.
2 Press NEW softkey.
3 Enter 14-digit serial number of the new system module.
4 The added system module appears in the list with the New status.
5 Save configuration change with ENTER or discard with Back.
Add a new I/O module
1 Select Set system modules menu item.
The list of the system modules available in the system is displayed.
2 Select the I/O module added, and automatically identified by the ana-
lyzer system, and press the NEW softkey.
3 When setting a digital I/O module:
Press FB-APPL. softkey and select function block applications.
4 The added system module appears in the list with the New status.
5 Save configuration change with ENTER or discard with Back.
NOTICE
If a Profibus module is refitted, it needs to be installed as the bottom I/O
module, i.e. on the slot -X20/-X21.