98 ACF5000 FTIR ANALYZER SYSTEM | OI/ACF5000-EN REV. A
Start Automatic Adjustment Check (AAC) manually
Function
The automatic cyclic monitoring of the adjustment of the FTIR spectrometer
(Automatic Adjustment Check – AAC) can be started manually. The handling
depends upon configuration of the AAC cycle to “split” or “complete” or
“complete & reference” (see page 78).
Menu path
Menu Adjust Automatic Adjustment FTIR
AAC cycle "split"
Display
Operation
START STEP
Start the next step in AAC cycle (see display).
AAC RESET
Reset step counter to 1.
STOP
Abort the running AAC step. The analyzer system is
purged with sample gas and returns to measuring
mode. No set values and actual values for the FTIR
span values are output. The AAC cannot be aborted
during automatic reference recording.