7-8 Troubleshooting and Global Customer Support 5702269-C
Advanced Energy
®
H004
Continued
from
previous
page
Continued
on next
page
Invalid Target
Selection
You have attempted to
select a target that is not
available.
Retry setting the target using
an acceptable value. See the
“I/O Communication and
Control” chapter for more
information.
Target Life Out
of Range
You have attempted to set
target life to a value outside
the allowable range.
Retry setting the target life
using an acceptable value.
See the command 12 in the
“I/O Communication and
Control” chapter for more
information.
Invalid Recipe
Step
You have attempted to enter
a recipe step that is invalid.
Retry setting the recipe step
using an acceptable value.
See the “I/O Communication
and Control” chapter for
more information.
Ramp Start
Point Out of
Range
You have attempted to set a
ramp start point to a value
outside the allowable range.
Re-adjust the ramp start
point by setting a smaller
percentage. See the “Setting
the Ramp” section in the “I/
O Communication and
Control” chapter for a
description of the ramp
option.
Table 7-3. Help Code Troubleshooting Table (Continued)
Help
Code
Number
Help Code
Message
Problem Indicated Suggested Action