341 Micro-cell Electron Capture Detector (µ-ECD)
Replacement procedures
8 of 26
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Figure 341-7 Removing the heater/sensor
4. Remove the two screws holding the upper heated block onto the assembly.
Carefully lift the block over the anode lead and remove.
5. Slide the heater and sensor out of the lower heated block.
Heater blocks
Heater/sensor assembly