EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6890 Series

Agilent Technologies 6890 Series Service Manual

Agilent Technologies 6890 Series
867 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #318 background imageLoading...
Page #318 background image
7 of 26
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Micro-cell Electron Capture Detector (µ-ECD) 341
Replacement procedures
WARNING The ECD micro-cell contains radioactive
63
Ni. To reduce the risk of exposure,
wear disposable gloves while handling the ECD micro-cell. When you are
finished, dispose of the gloves and wash your hands with soap and water.
1. Remove the detector as described in the Replacing the entire detector/
detector cell procedure in this section.
2. Loosen the locking tab screw on top of the detector, slide the locking tab
back, and pivot it out of the away.
Figure 341-6 Unlocking the thermal cover
3. Lift the thermal cover up and carefully slide it off of the anode assembly.

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals