Agilent 7820 GC Service
131
5
Performance Verification
About Chromatographic Checkout 131
To Prepare for Chromatographic Checkout 132
To Check FID Performance 132
To Check NPD Performance 136
To check TCD Performance 140
To Check uECD Performance 145
To Check FPD Performance (Sample 5188-5953)149
About Chromatographic Checkout
The tests described in this section provide basic confirmation
that the GC and detector can perform comparably to factory
condition. However, as detectors and the other parts of the GC
age, detector performance can change. The results presented
here represent typical outputs for typical operating conditions
and are not specifications.
The tests assume the following:
ï‚· Use of an automatic liquid sampler. If not available, use a
suitable manual syringe instead of the syringe listed.
 Use of a 10-µL syringe in most cases. However, a 5-µL
syringe is an acceptable substitute for the 1-µL injections
described here.
ï‚· Use of the septa and other hardware (liners, jets, adapters,
and so forth) described. If you substitute other hardware,
performance can vary.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals