EasyManuals Logo

Agilent Technologies 7820A Service Manual

Agilent Technologies 7820A
397 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
24
Agilent 7820 GC Service
Inlet overview
Inlets are a means of introducing a sample into the carrier gas
stream and then into the analyzing column. The term "Inlets"
includes:
ï‚§ Injection ports, for use with a syringe
ï‚§ Sampling valves for high sample size reproducibility
Split/Splitless Inlet
This is the general-purpose inlet for use with capillary columns.
It vaporizes the sample in a heated liner, then delivers all
(splitless mode) or a specified fraction (split mode) of the vapor
to the column.
Replacing the split/splitless inlet
Cool down the oven.
Turn off the GC main power switch and disconnect its power
cord.
Hazardous voltages are present in the
mainframe when the GC power cord is connected.
Avoid a potentially dangerous shock hazard by
disconnecting the power cord before removing
any GC panels.
Components can be damaged by static electricity:
be sure to wear an ESD strap grounded to the GC
chassis while performing this procedure.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals