Converting the power option
This procedure outlines the steps you must complete to prepare
a GC for use in another country or to configure for a different
oven heating speed.
Procedure
1 Select the appropriate power option kit. See Power options (
257). Compare the parts listings for the present and new
power option to see what must be replaced.
Select the appropriate power cord. See Power cords ( 396).
Cool down the oven.
Turn off the GC main power switch and disconnect its power
cord.
Hazardous voltages are present in the
mainframe when the GC power cord is connected.
Avoid a potentially dangerous shock hazard by
disconnecting the power cord before removing
any GC panels.
Components can be damaged by static electricity:
be sure to wear an ESD strap grounded to the GC
chassis while performing this procedure.
If necessary, replace the oven shroud. See Replacing the oven
shroud, oven fan, and oven fan motor ( 205).

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals