Agilent 7820 GC Service
261
Remote start/stop connection
Prepare (low)—Request to prepare for analysis. Receiver is any
module performing pre-analysis activities.
Start (low)—Request to start run/timetable. Receiver is any
module performing runtime-controlled activities.
Ready (high)—System is ready for next analysis. Receiver is
any sequence controller.
Stop (low)—Request to reach system ready state as soon as
possible (for example, stop run, abort or finish, and stop
injection). Receiver is any module performing runtime-controlled
activities.

Table of Contents