Agilent 7820 GC Service
Zero the pressure sensors.
Configure the gas types and control mode.
a Press [Config], then scroll to the correct PCM entry (PCM
A or PCM B).
b Scroll to Gas type and use the [Mode/Type] key to set
the gas type connected to the PCM's Channel 1.
Configuring time
To localize the date/time stamp in the Run and Event logs, use
this procedure.
1 Press [Config][Time].
Enter the offset from Greenwich Mean Time using 24-hour
format. Press [Enter].
Enter the local time (24 hour format). Press [Enter].
Enter the date, in ddmmyy format. Press [Enter].