Agilent 7820 GC Service
145
output, Range 0)
>>Set DATE and TIME<<
Zero
10
Pk wd
0.04
Att^2
7
Thresh
5
Cht sp
1
Stop time
8.75
Ar Rej
1000
SS
PP
Sensitivity
C , C , C area counts
525000
460000
Discrimination
C /C area ratio
1.00 +/- 0.10
1.00 +/- 0.10
To Check uECD Performance
1 Gather the following:
 Evaluation column, HP-5 30 m — 0.32 mm — 0.25 µm
(19091J-413)
ï‚§ uECD performance evaluation (checkout) sample
(18713-60040, Japan: 5183-0379)
ï‚§ 4-mL solvent and waste bottles or equivalent for
autoinjector
ï‚§ Chromatographic-grade isooctane
ï‚§ 2-mL sample vials or equivalent for sample
Verify the following:
ï‚§ Clean fused silica indented mixing liner installed. If not,
install it.
ï‚§ Chromatographic-grade gases plumbed and configured:
helium as carrier, nitrogen as makeup. (Nitrogen is an
acceptable alternate carrier gas.)
ï‚§ Empty waste vials loaded in sample turret.
ï‚§ 4-mL vial with diffusion cap filled with hexane and inserted
in Solvent A injector position.
Install the evaluation column. (See the procedure for the SS or
PP in the Maintenance manual.)
ï‚§ Bake out the evaluation column for at least 30 minutes at

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals