EasyManua.ls Logo

Agilent Technologies 7820A

Agilent Technologies 7820A
397 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
144
Agilent 7820 GC Service
The following are typical values for the detector results. The
results for a detector depend on site preparation related factors.
These are not pass/fail criteria.
Agilent ChemStation
SS
PP
Sensitivity
C , C , C area counts
65
58
Discrimination
C /C area ratio
1.00 +/- 0.10
1.00 +/- 0.10
Agilent ChemStation ASTM noise
Measured with GC isothermal at 100 °C
10 minute blank run, noise range(s) > 1 minute
Performance + noise report
TCD ASTM noise 0.064 display units (25 µV/display unit)
Detector signal set to 5 Hertz
3396B or 3396C or 3397A integrator sample
conditions with analog input (GC Analog

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals