Agilent 7820 GC Service
65
Nitrogen Phosphorus Detector (NPD)
The NPD is a variation on the FID, in that the sample is burned
and the resulting ions are collected. However, the hydrogen/air
ratio is set to suppress carbon ionization and an alkali salt
enhances nitrogen and phosphorus ionization.
Selecting an NPD jet
Open the oven door and locate the column connection fitting at
the base of the detector. It will look like either a capillary
optimized fitting or an adaptable fitting.
ï‚· If you have an application that tends to clog the jet, select a
jet with a wider tip id.
ï‚· When using packed columns in high column-bleed
applications, the jet tends to clog with silicon dioxide.
For adaptable fittings, select one of the following from the table
below, "Jets for adaptable fittings".
Table 37 Jets for adaptable fittings
Figure
4 ID
Jet type
Part number
Jet tip id
Length
1
Capillary with
extended jet
(recommended)
G1534-80590
0.29 mm
(0.11
inch)
70.5
mm
2
Capillary
19244-80560
0.29 mm
(0.011
inch)
61.5
mm

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals