Agilent 7820 GC Service
Push the wire brush through the collector.
Rinse the parts in tap water or distilled water.
Rinse all surfaces of the parts in methanol; inside and outside
surfaces. To insure good rinsing use either a Teflon wash bottle
or a beaker.
From this point on, handle the parts only with forceps (or
tweezers). Remove the jet from the bath and rinse it thoroughly
with hot tap water and then with a small amount of methanol.
Blow the jet dry with a burst of compressed air or nitrogen and
then place the jet on a clean cloth to air dry.