EasyManuals Logo

Agilent Technologies 7820A Service Manual

Agilent Technologies 7820A
397 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
64
Agilent 7820 GC Service
Push the wire brush through the collector.
Rinse the parts in tap water or distilled water.
Rinse all surfaces of the parts in methanol; inside and outside
surfaces. To insure good rinsing use either a Teflon wash bottle
or a beaker.
From this point on, handle the parts only with forceps (or
tweezers). Remove the jet from the bath and rinse it thoroughly
with hot tap water and then with a small amount of methanol.
Blow the jet dry with a burst of compressed air or nitrogen and
then place the jet on a clean cloth to air dry.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals