Agilent 7820 GC Service
263
Temperature sensor resistance
The approximate resistance of a temperature sensor is:
R = 100 + (.35 ï‚´ t)
Where R is resistance in ohms and t is sensor temperature in °C.
Temperature Sensor Resistance by Heater Temperature
°C
+0°
+10°
+20°
+30°
+40°
+50°
+60°
+70°
+80°
+90°
+100°
0°
100.00
103.90
107.79
111.67
115.54
119.40
123.24
127.07
130.89
134.70
138.50
100°
138.50
142.28
146.06
149.82
153.57
157.32
161.04
164.76
168.47
172.16
175.84
200°
175.84
179.51
183.17
186.82
190.46
194.08
197.70
201.30
204.88
208.46
212.03
300°
212.03
215.58
219.13
222.66
226.18
229.69
233.19
236.67
240.15
243.61
247.06
400°
247.06
250.50
253.34
257.34
260.75
264.14
267.52
270.89
274.25
277.60
280.93
Ceramic and glass fuses
There are two sets of fuses on the AC power board:
ï‚· Two ceramic fuses protect the board from the oven heating
circuit (F3, F4)
ï‚· Two glass fuses protect the board from the main supply (F1,
F2)
See the table below for the part numbers. When replacing both
the glass and ceramic fuse types, always replace them in pairs.
I.D
.
Description
System
Part
number

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals