Agilent 7820 GC Service
Purge time
0.5 min
Gas saver
Off
Purged packed
inlet
Temperature
250 °C
Detector
Temperature
300 °C
H2 flow
30 mL/min
Air flow
400 mL/min
Makeup flow (N2)
25 mL/min
Lit offset
Typically 2 pA
Oven
Initial temp
75 °C
Initial time
0.5 min
Rate 1
20 °C/min
Final temp
190 °C
Final time
0 min
ALS settings (if installed)
Sample washes
2
Sample pumps
6
Sample wash volume
8
Injection volume
1 µL
Syringe size
10 µL
Solvent A pre
washes
2
Solvent A post
washes
2
Solvent A wash
volume
8
Solvent B pre
washes
0
Solvent B post
washes
0
Solvent B wash
volume
0
Injection mode
Normal
Airgap volume
0.20
Viscosity delay
0 s
Inject Dispense
Speed
6000
PreInjection dwell
0
PostInjection
dwell
0
Manual injection

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals