EasyManua.ls Logo

Agilent Technologies 7820A

Agilent Technologies 7820A
397 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
134
Agilent 7820 GC Service
Purge time
0.5 min
Gas saver
Off
Purged packed
inlet
Temperature
250 °C
Detector
Temperature
300 °C
H2 flow
30 mL/min
Air flow
400 mL/min
Makeup flow (N2)
25 mL/min
Lit offset
Typically 2 pA
Oven
Initial temp
75 °C
Initial time
0.5 min
Rate 1
20 °C/min
Final temp
190 °C
Final time
0 min
ALS settings (if installed)
Sample washes
2
Sample pumps
6
Sample wash volume
8
Injection volume
1 µL
Syringe size
10 µL
Solvent A pre
washes
2
Solvent A post
washes
2
Solvent A wash
volume
8
Solvent B pre
washes
0
Solvent B post
washes
0
Solvent B wash
volume
0
Injection mode
Normal
Airgap volume
0.20
Viscosity delay
0 s
Inject Dispense
Speed
6000
PreInjection dwell
0
PostInjection
dwell
0
Manual injection

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals