EasyManuals Logo

Agilent Technologies 7820A Service Manual

Agilent Technologies 7820A
397 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
Agilent 7820 GC Service
137
(19091J-413)
ï‚§ NPD performance evaluation (checkout) sample
(18789-60060)
ï‚§ 4-mL solvent and waste bottles or equivalent for
autoinjector
ï‚§ Chromatographic-grade isooctane
ï‚§ 2-mL sample vials or equivalent for sample
Verify the following:
ï‚§ Capillary column jet installed. If not, select and install a
capillary column jet.
ï‚§ Capillary column adapter installed. If not, install it.
ï‚§ Chromatographic-grade gases plumbed and configured:
helium as carrier gas, nitrogen, hydrogen, and air.
ï‚§ Empty waste vials loaded in sample turret.
ï‚§ 4-mL vial with diffusion cap filled with isooctane and
inserted in Solvent A injector position.
If present, remove any protective caps from the inlet manifold
vents.
Install the evaluation column. (See the procedure for the SS or
PP in the Maintenance manual.)
ï‚§ Bake out the evaluation column for at least 30 min at
180 °C. (See the procedure for the SS or PP in the
Maintenance manual.)
ï‚§ Be sure to configure the column.
Create or load a method with these parameter values.
Table 2. NPD Checkout Conditions
Column and sample
Type
HP-5, 30 m — 0.32 mm — 0.25 µm
(19091J-413)
Sample
NPD checkout 18789-60060
Column mode
Constant flow
Column pressure
6.5 mL/min (helium)
Split/splitless
inlet
Temperature
200 °C
Mode
Splitless
Purge flow
60 mL/min

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals