Agilent 7820 GC Service
141
2-mL sample vials or equivalent for sample
Chromatographic-grade helium as carrier, makeup, and
reference gas
Verify the following:
Chromatographic-grade gases plumbed and configured:
helium as carrier gas, nitrogen, hydrogen, and air.
Empty waste vials loaded in sample turret.
4-mL solvent vial with diffusion cap filled with isooctane
and inserted in Solvent A injector position.
Install the evaluation column. (See the procedure for the SS or
PP in the Maintenance manual.)
Bake out the evaluation column for at least 30 min at
180 °C. (See the procedure for the SS or PP in the
Maintenance manual.)
Be sure to configure the column.
Create or load a method with these parameter values.
TCD Checkout Conditions
HP-5, 30 m — 0.32 mm — 0.25 µm
(19091J-413)
FID/TCD checkout 18710-60170
< 30 display counts on Agilent
ChemStation (< 750 µV)