Agilent 7820 GC Service
141
ï‚§ 2-mL sample vials or equivalent for sample
ï‚§ Chromatographic-grade helium as carrier, makeup, and
reference gas
Verify the following:
ï‚§ Chromatographic-grade gases plumbed and configured:
helium as carrier gas, nitrogen, hydrogen, and air.
ï‚§ Empty waste vials loaded in sample turret.
ï‚§ 4-mL solvent vial with diffusion cap filled with isooctane
and inserted in Solvent A injector position.
Install the evaluation column. (See the procedure for the SS or
PP in the Maintenance manual.)
ï‚§ Bake out the evaluation column for at least 30 min at
180 °C. (See the procedure for the SS or PP in the
Maintenance manual.)
ï‚§ Be sure to configure the column.
Create or load a method with these parameter values.
TCD Checkout Conditions
Column and sample
Type
HP-5, 30 m — 0.32 mm — 0.25 µm
(19091J-413)
Sample
FID/TCD checkout 18710-60170
Column flow
6.5 mL/min
Column mode
Constant flow
Split/splitless
inlet
Temperature
250 °C
Mode
Splitless
Purge flow
60 mL/min
Purge time
0.75 min
Purged packed
inlet
Temperature
250 °C
Detector
Temperature
300 °C
Reference flow
(He)
20 mL/min
Makeup flow (He)
2 mL/min
Baseline output
< 30 display counts on Agilent
ChemStation (< 750 µV)
Oven
Initial temp
40 °C

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals