148
Agilent 7820 GC Service
If using a data system, prepare the system to perform one run
using the loaded checkout method. Make sure that the data
system will output a chromatogram.
Start the run.
If performing an injection using an autosampler, start the run
using the data system or press [Start] on the GC.
If performing a manual injection (with or without a data
system):
a Press [Prep Run] to prepare the inlet for splitless
injection.
b When the GC becomes ready, inject 1 µL of the checkout
sample and press [Start] on the GC.
The following chromatogram shows typical results for a new
detector with new consumable parts installed.