Agilent 7820 GC Service
227
room to pass through the largest of the three cable connectors.
Note the orientation of the old transformer with respect to the
various cables, then remove the Allen bolt to release the
transformer. Carefully lift it from the GC.
In replacement, make sure the lower pad is in place, then orient
the new transformer onto the pad.
Replace the upper pad and associated hardware. In restoring
cables, feed the large cable and connector through to the
electronics side of the GC first.