EasyManua.ls Logo

Agilent Technologies 7820A

Agilent Technologies 7820A
397 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
88
Agilent 7820 GC Service
It may take 24 hours for the uECD baseline to
completely stabilize, especially if you are
starting with a cold system and want to assure
high-sensitivity operation. Therefore, for the
most accurate results, run the detector at normal
operating conditions for as long as possible (at
least 2 hours and up to 24 hours) before running
the frequency test.
If you will be injecting into an unused inlet, you
must use low-bleed septa. Make sure to condition
new septa before use in an inlet for several hours
with 1 to 5 mL/min carrier flow.
1 Make sure you are using normal operating conditions and
that at least 2 hours have elapsed since the last run.
Turn on the uECD and the corresponding signal.
Check the displayed "Output":
<25 = uECD frequency is acceptable
1000 = Contaminants in system
Each display count equals a frequency of 1 Hertz
(e.g., a display reading of 100 = 100 Hz.).
If the uECD frequency indicates contamination (1000) check for
the following:
Contaminated carrier gas trap(s) and or supplyreplace
carrier gas supply tank and any traps on the carrier supply
line.
Insufficient column conditioningfully condition the
column.
Contaminated detectorbake out the detector.
Column, inlet and/or septum bleedclean the
inlet/replace the septum with a conditioned, low bleed
septum.
Leaksperform leak tests on both the inlet and detector
systems.
Anode current leakagemake sure the anode contacts
are clean. Make sure the anode nut is tight.

Table of Contents

Other manuals for Agilent Technologies 7820A

Related product manuals