MMEMory:LOAD:STATe <filename>
MMEMory:STORe:STATe <filename>
Loads an instrument state from a state file or saves the current instrument state to a state file. The file name
optionally includes the folder name and the .sta file extension. If you do not include this extension, the instrument
firmware will automatically append it.
Parameter Typical Return
Any valid file name (none)
Store the current instrument state to the specified state file in the root directory of the internal flash file system:
MMEM:STOR:STAT "INT:\MySetup"
Load the instrument state from MySetup.sta saved by the previous command:
MMEM:LOAD:STAT "INT:\MySetup.sta"
l The specified folder must exist and cannot be marked as hidden or system.
l The specified file cannot be empty or marked as hidden or system.
l The state file created by *SAV 0, called STATE_0.sta, resides in the root folder of the internal flash file system. It is
overwritten with the power-down state of the instrument when power is cycled.
MMEMory:STATe:RECall:AUTO {ON|1|OFF|0}
MMEMory:STATe:RECall:AUTO?
Disables or enables the automatic recall of a specific stored instrument state at power-on. Specify ON to automatically
recall the power-down state file (STATE_0 in the root folder of the internal flash file system) or a user-specified state file
(MMEMory:STATe:RECall:SELect). Select "OFF" to issue a Factory Reset (*RST) at power-on.
Parameter Typical Return
{ON|1|OFF|0}, default ON 0(OFF)or1(ON)
Select the power-down state to be used at power-on.
MMEM:STAT:REC:SEL "INT:\STATE_0"
MMEM:STAT:REC:AUTO ON
Save the current state into a file in the root directory of the internal flash file system and load it at power-on:
MMEM:STOR:STAT "INT:\MyVoltMeas"
MMEM:STAT:REC:SEL "INT:\MyVoltMeas"
MMEM:STAT:REC:AUTO ON
l When shipped from the factory, the instrument is configured to automatically recall the power-down state file
when power is restored.
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
Agilent Truevolt Series DMM Operating and Service Guide 181

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals