[SENSe:]{FREQuency|PERiod}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]{FREQuency|PERiod}:NULL[:STATe]?
Enables or disables the null function for frequency and period measurements.
Unlike the SENSe:FREQuency and SENSe:PERiod range and aperture commands, this parameter is not
shared between frequency and period measurements. The null parameters are independent for
frequency and period measurements.
Parameter Typical Return
{ON|1|OFF|0}, default OFF 0ξ˜ƒ(OFF)ξ˜ƒorξ˜ƒ1ξ˜ƒ(ON)
Configure frequency measurements, using the null function to subtract 1kHz from the measurements. Make and
read two measurements:
CONF:FREQ
FREQ:NULL:STAT ON;VAL 1 kHz
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+03,+1.04570000E+03
l Enabling the scaling function also enables automatic null value selection ([SENSe:]
{FREQuency|PERiod}:NULL:VALue:AUTO ON).
l To set a fixed null value, use this command: [SENSe:]{FREQuency|PERiod}:NULL:VALue.
l The instrument disables the null function after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or
CONFigure function.
[SENSe:]{FREQuency|PERiod}:NULL:VALue {<value>|MIN|MAX|DEF}
[SENSe:]{FREQuency|PERiod}:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for frequency and period measurements.
Unlike the SENSe:FREQuency and SENSe:PERiod range and aperture commands, this parameter is not
shared between frequency and period measurements. The null parameters are independent for
frequency and period measurements.
Parameter Typical Return
-1.2E6 to +1.2E6, default 0 +1.00000000E-02
Configure frequency measurements, using the null function to subtract 1kHz from the measurements. Make and
read two measurements:
CONF:FREQ
FREQ:NULL:STAT ON;VAL 1 kHz
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+03,+1.04570000E+03
l Specifying a null value disables automatic null value selection ([SENSe:]{FREQuency|PERiod}:NULL:VALue:AUTO
OFF).
l To use the null value, the null state must be on ([SENSe:]{FREQuency|PERiod}:NULL:STATe ON).
l This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or
CONFigure function.
Agilent Truevolt Series DMM Operating and Service Guide 199

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals