[SENSe:]{RESistance|FRESistance}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]{RESistance|FRESistance}:NULL[:STATe]?
Enables or disables the null function for all resistance measurements.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter Typical Return
{ON|1|OFF|0}, default OFF 0(OFF)or1(ON)
Configure 2-wire resistance measurements, using the null function to remove 100 mΩ of wiring resistance. Make and
read two measurements:
CONF:RES
RES:NULL:STAT ON;VAL .1
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
l Enabling the scaling function also enables automatic null value selection (SENSe:
{RESistance/FRESistance}:NULL:VALue:AUTO ON).
l To set a fixed null value, use this command: [SENSe:]{RESistance/FRESistance}:NULL:VALue.
l The instrument disables the null function after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or
CONFigure function.
[SENSe:]{RESistance|FRESistance}:NULL:VALue {<value>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for all resistance measurements.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter Typical Return
-120 to +120 MΩ, default 0 +1.04530000E+02
Configure 2-wire resistance measurements, using the null function to remove 100 mΩ of wiring resistance. Make and
read two measurements:
CONF:RES
RES:NULL:STAT ON;VAL .1
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
l Specifying a null value disables automatic null value selection ([SENSe:]
{RESistance/FRESistance}:NULL:VALue:AUTO OFF).
l To use the null value, the null state must be on ([SENSe:]{RESistance|FRESistance}:NULL:STATe ON).
l This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or
CONFigure function.
Agilent Truevolt Series DMM Operating and Service Guide 205

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals