[SENSe:]TEMPerature:NULL:VALue:AUTO {ON|1|OFF|0}
[SENSe:]TEMPerature:NULL:VALue:AUTO?
Enables or disables automatic null value selection for temperature measurements.
Parameter Typical Return
{ON|1|OFF|0}, default OFF 0(OFF)or1(ON)
Configure temperature measurements, using the null function to subtract 25° from them. Make and read two
measurements:
CONF:TEMP RTD
TEMP:NULL:STAT ON;VAL 25>
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+00,+1.04570000E+00
Make a second set of measurements using automatic null value selection:
TEMP:NULL:VAL:AUTO ON
READ?
Typical Response: +0.00000000E+00,+0.01420000E+00
l When automatic reference selection is ON, the first measurement made is used as the null value for all
subsequent measurements. [SENSe:]TEMPerature:NULL:VALue will be set to this value. Automatic null value
selection will be disabled.
l When automatic null value selection is disabled (OFF), the null value is specified by this command: [SENSe:]
TEMPerature:NULL:VALue.
l The instrument enables automatic null value selection when the null function is enabled ([SENSe:]
TEMPerature:NULL:STATe ON).
l This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset (SYSTem:PRESet), or
CONFigure function.
[SENSe:]TEMPerature:TRANsducer:{RTD|FRTD}:RESistance[:REFerence]
{<reference>|MIN|MAX|DEF}
[SENSe:]TEMPerature:TRANsducer:{RTD|FRTD}:RESistance[:REFerence]?
[{MIN|MAX|DEF}]
Selects the nominal resistance (R
0
) for 2-wire and 4-wire RTD measurements. R
0
is the nominal resistance of an RTD
at 0 °C.
Parameter Typical Return
80 Ω to 120 Ω, default 100 Ω +1.00100000E+02
Configure 4-wire RTD measurements using an RTD with an R
0
of 100.1 Ω. Make and read two measurements:
CONF:TEMP FRTD
TEMP:TRAN:FRTD:RES 100.1
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
Agilent Truevolt Series DMM Operating and Service Guide 213

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals