[SENSe:]TEMPerature:TRANsducer:{THERmistor|FTHermistor}:TYPE 5000
[SENSe:]TEMPerature:TRANsducer:{THERmistor|FTHermistor}:TYPE?
Selects the thermistor type for 2-wire and 4-wire temperature measurements. Choosing the THERmistor or the
FTHermistor command form makes no difference, as both commands affect the same parameter.
Parameter Typical Return
5000 (only available choice) +5000
Configure 4-wire thermistor measurements. Make and read two measurements:
CONF:TEMP FTH
TEMP:TRAN:FTH:TYPE 5000
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
[SENSe:]TEMPerature:TRANsducer:TYPE {FRTD|RTD|FTHermistor|THERmistor}
[SENSe:]TEMPerature:TRANsducer:TYPE?
Selects the transducer probe type to use for temperature measurements. The supported probes are 2-wire and 4-wire
RTDs and 2-wire and 4-wire thermistors.
Parameter Typical Return
{FRTD|RTD|FTHermistor|THERmistor}, default FRTD FRTD, RTD, FTH, or THER
Configure 2-wire RTD measurements using an RTD with an R
0
of 100.1 Ω. Make and read two measurements:
FUNC "TEMP"
TEMP:TRAN:TYPE RTD
TEMP:TRAN:FRTD:RES 100.1
SAMP:COUN 2
READ?
Typical Response: +2.54530000E+01,+2.54570000E+01
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
214 Agilent Truevolt Series DMM Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals