[SENSe:]TEMPerature:TRANsducer:{THERmistor|FTHermistor}:TYPE 5000
[SENSe:]TEMPerature:TRANsducer:{THERmistor|FTHermistor}:TYPE?
Selects the thermistor type for 2-wire and 4-wire temperature measurements. Choosing the THERmistor or the
FTHermistor command form makes no difference, as both commands affect the same parameter.
Parameter Typical Return
5000 (only available choice) +5000
Configure 4-wire thermistor measurements. Make and read two measurements:
CONF:TEMP FTH
TEMP:TRAN:FTH:TYPE 5000
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
[SENSe:]TEMPerature:TRANsducer:TYPE {FRTD|RTD|FTHermistor|THERmistor}
[SENSe:]TEMPerature:TRANsducer:TYPE?
Selects the transducer probe type to use for temperature measurements. The supported probes are 2-wire and 4-wire
RTDs and 2-wire and 4-wire thermistors.
Parameter Typical Return
{FRTD|RTD|FTHermistor|THERmistor}, default FRTD FRTD, RTD, FTH, or THER
Configure 2-wire RTD measurements using an RTD with an R
0
of 100.1 Ω. Make and read two measurements:
FUNC "TEMP"
TEMP:TRAN:TYPE RTD
TEMP:TRAN:FRTD:RES 100.1
SAMP:COUN 2
READ?
Typical Response: +2.54530000E+01,+2.54570000E+01
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
214 Agilent Truevolt Series DMM Operating and Service Guide