SYSTem:COMMunicate:LAN:TELNet:PROMpt "<string>"
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt seen when communicating with the instrument via Telnet.
Parameter Typical Return
String of up to 15 characters
Default: 34460A> or 34461A>
"Command>"
Set the command prompt:
SYST:COMM:LAN:TELN:PROM "Command>"
l On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l The instrument uses LAN port 5024 for SCPI Telnet sessions and port 5025 for SCPI Socket sessions.
l Telnet sessions are typically started from a host computer shell:
telnet <IP_address> <port>
For example:
telnet 169.254.4.10 5024
To exit a Telnet session, press <Ctrl-D>.
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
l SYSTem:SECurity:IMMediate sets this parameter to its default value.
SYSTem:COMMunicate:LAN:TELNet:WMESsage "<string>"
SYSTem:COMMunicate:LAN:TELNet:WMESsage?
Sets welcome message seen when communicating with instrument via Telnet.
Parameter Typical Return
Quoted string of up to 63 characters.
Default: "Welcome to Agilent's 34460A Digital Multimeter" or "Welcome to
Agilent's 34461A Digital Multimeter"
"Welcome to the Telnet
Session"
Define a welcome message:
SYST:COMM:LAN:TELN:WMES "Welcome to the Telnet Session"
l On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l The instrument uses LAN port 5024 for SCPI Telnet sessions and port 5025 for SCPI Socket sessions.
l This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instrument Preset
(SYSTem:PRESet).
l SYSTem:SECurity:IMMediate sets this parameter to its default value.
246 Agilent Truevolt Series DMM Operating and Service Guide