SYSTem:LOCK:OWNer?
Returns the I/O interface that currently has a lock.
Parameter Typical Return
(none) "LAN169.254.149.35"
See Interface Locking Example
l When a lock is active, Bit 10 in the Standard Operation Register will be set (STATus:OPERation:CONDition?). When
the lock is released on all I/O interfaces, this bit will be cleared.
l Returns "USB", "VXI11", "GPIB", or "LAN <IP Address>" indicating the I/O interface that currently has a lock. If
no interface has a lock, "NONE" is returned.
SYSTem:LOCK:RELease
Decrements the lock count by 1 and may release the I/O interface from which the command is executed.
Parameter Typical Return
(none) (none)
See Interface Locking Example
l When a lock is active, Bit 10 in the Standard Operation Register will be set (STATus:OPERation:CONDition?). When
the lock is released on all I/O interfaces, this bit will be cleared.
SYSTem:LOCK:REQuest?
Requests a lock of the current I/O interface. This allows you to lock the instrument's configuration or cooperatively
share the instrument with other computers.
Parameter Typical Return
(none) +0 (denied) or +1 (granted)
See Interface Locking Example
l Lock requests can be nested; each request increases lock count by 1. For every request, you will need a release
(SYSTem:LOCK:RELease) from the same I/O interface.
l Locks are handled at the I/O interface level (USB, LAN, etc.) and you are responsible for all coordination between
threads and/or programs on that interface.
l When a request is granted, only I/O sessions from the present interface can change the instrument state. You can
only query the instrument state from other I/O interfaces.
l LAN sessions locks are automatically released when a LAN disconnect is detected.
l Granting a lock sets Bit 10 in the Standard Operation Register (STATus:OPERation:CONDition?). In addition, the
entire instrument front panel, including the Local key, is locked out while a lock is in place ("Front panel locked." is
displayed).
250 Agilent Truevolt Series DMM Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals