Fault removal
multi EA 4000 Version 01.16 105
14 Data transfer interrupted
no data transfer for 10 s
− initialize
17 Incorrect interface protocol ID
Cause Remedy
error following update
(program versions firmware – mul-
tiWin
®
do not match)
− update required
20 Timeout: Intend
Cause Remedy
timeout during initialization
− initialize
21 Timeout: Status Busy
timeout during operation (device
not ready for measurement)
− acknowledge message
− initialize
22 Timeout: End
timeout when exiting multiWin
− acknowledge message
− initialize
23 Timeout: Stop End
Cause Remedy
timeout during measurement can-
cellation
− acknowledge message
− initialize
24
Cause Remedy
timeout during measurement
maximum integration time exceed-
ed
− check analyzer
− check method configuration
25 Timeout: Analyzer command processing
timeout during command pro-
cessing
− acknowledge message
− initialize
30 Chloride needs adding
chloride needs adding
− open menu item S
YSTEM /
COMPONENT TEST
50
Cause Remedy
internal computer (firmware) has
restarted.
− acknowledge message
− initialize
52
Cause Remedy
firmware was run in standby
− acknowledge message
− initialize