Preventive Maintenance
Progressa+ Bed Instructions for Use (216321 REV 2) 103
PREVENTIVE MAINTENANCE
Warning—To help prevent injury and/or equipment damage, obey these warnings:
• Warning—Only facility-authorized personnel should perform calibration on the Progressa+
Bed.
• Warning—Only facility-authorized personnel should perform preventive maintenance on the
Progressa+ Bed.
The Progressa+ Bed requires an effective maintenance program. We recommend that you perform
annual preventive maintenance (PM) and testing for Joint Commission certification. PM and testing not
only meet Joint Commission requirements but can help support a long, operative life for the
Progressa+ Bed. PM will minimize downtime due to excessive wear. For the preventive maintenance
detailed procedures, refer to the Progressa and Progressa+ Bed Service Manual (171748).
Perform annual preventive maintenance procedures to make sure all bed components are functioning
as originally designed. Pay particular attention to safety features, including but not limited to:
• Overall condition
• Siderail latching mechanisms, dampers and deck
• Brakes, Steer and casters
• IntelliDrive Transport System (if applicable)
•Batteries
• Scale system
• Motors, frame movement and head angle display
• AC power cords for fraying, damage, proper grounding and electrical safety
• All controls return to off or neutral position when released
• Controls or cabling entanglement in system mechanisms or siderails
• Proper operation of the lockout controls
• Surface, top cover and blower
• Intake/Exhaust filters
•SideCom Communication System (if applicable)
• Accessories