EasyManua.ls Logo

baxter Hillrom Proqressa+ - Preventive Maintenance

baxter Hillrom Proqressa+
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preventive Maintenance
Progressa+ Bed Instructions for Use (216321 REV 2) 103
PREVENTIVE MAINTENANCE
WARNING:
WarningTo help prevent injury and/or equipment damage, obey these warnings:
Warning—Only facility-authorized personnel should perform calibration on the Progressa+
Bed.
Warning—Only facility-authorized personnel should perform preventive maintenance on the
Progressa+ Bed.
The Progressa+ Bed requires an effective maintenance program. We recommend that you perform
annual preventive maintenance (PM) and testing for Joint Commission certification. PM and testing not
only meet Joint Commission requirements but can help support a long, operative life for the
Progressa+ Bed. PM will minimize downtime due to excessive wear. For the preventive maintenance
detailed procedures, refer to the Progressa and Progressa+ Bed Service Manual (171748).
Perform annual preventive maintenance procedures to make sure all bed components are functioning
as originally designed. Pay particular attention to safety features, including but not limited to:
Overall condition
Siderail latching mechanisms, dampers and deck
Brakes, Steer and casters
IntelliDrive Transport System (if applicable)
•Batteries
Scale system
Motors, frame movement and head angle display
AC power cords for fraying, damage, proper grounding and electrical safety
All controls return to off or neutral position when released
Controls or cabling entanglement in system mechanisms or siderails
Proper operation of the lockout controls
Surface, top cover and blower
Intake/Exhaust filters
•SideCom Communication System (if applicable)
Accessories