Alerts and Information Indicators
14 Progressa+ Bed Instructions for Use (216321 REV 2)
ALERTS AND INFORMATION INDICATORS
The Information Indicators provide the caregiver with visual indications about: Audible Indicators,
Battery Status, Service Required, Hip Position Locator, and head section angle.
AUDIBLE INDICATORS
A single beep will sound when an activity is successful.
A triple beep will sound when to indicate that you should look at the GCI screen, attention is needed.
BED BATTERY POWER
CAUTION:
Caution—Although a fully-charged battery is preferred, transport may be done when the battery
charge is low. The bed should be reconnected to AC power as soon as possible to prevent equipment
damage.
If the Battery Indicator changes from Charged to Low consistently within four hours of being
disconnected from AC power, replace the battery.
While on battery power, the bed will operate as follows:
All bed articulations will operate
Integrated surfaces will remain inflated, but will not adjust pressures
The GCI screen will not display
SERVICE REQUIRED
Charged - The Charged indicator (+) comes on when the battery is
charged.
Low - The Low indicator (-) flashes when the battery is low. An
intermittent tone sounds every two minutes when the battery
reaches low condition and AC is unplugged.
Off - If the battery is too low to operate the bed.
NOTE:
If the bed is unplugged, press any function to activate the battery
power status.
The Service required indicator illuminates when the bed detects a
malfunction. Contact the facility-authorized maintenance persons
or local Hill-Rom representative for assistance.