Bed Frame Features
88 Progressa+ Bed Instructions for Use (216321 REV 2)
COMPOSER COMMUNICATION SYSTEM
The Progressa+ Bed is compatible with the COMposer Communication System. With the COMposer
Communication System, the bed can be monitored for the following functions:
•Bed in Low position
• Siderail(s) up or down
•Brake set
• Bed exit on or off
VOALTE NURSE CALL SYSTEM
The Voalte Nurse Call System is an enterprise system that connects and monitors Baxter and Hill-Rom
beds and surfaces. The system sends bed and surface data to network applications for caregivers to
view and receive alerts. For complete operational instructions for the Voalte Nurse Call System refer to
the Voalte Nurse Call System User Manual.
WIRELESS CONNECTIVITY
Warning—The wireless module does not communicate nurse call information. The bed’s SideCom
Communication System cable must be connected to the facility network for remote nurse call
communications. Failure to connect the SideCom Communication System cable may result in delay of
critical care to the patient.
The Wireless Connectivity modules are not intended as a replacement for your wired Nurse Call
connection.
This module does not provide wireless use of environmental controls such as audio or room lighting.
The module operates only when the bed is connected to AC power; it does not operate on battery
power.
This module is compliant with 2014/53/EU—Radio Equipment Directive (RED).
Module Location
To determine if the bed has the wireless module
installed, look at the left foot-end lift arm. If there is a
wireless antenna mounted on the left foot-end lift
arm, the bed has the wireless module installed.