Technical Specifications
116 Progressa+ Bed Instructions for Use (216321 REV 2)
Applied Parts (in accordance with IEC 60601-1)
Scale Classification (European Scale Beds Only)
Nurse Call Connection Requirements
For information about the Nurse Call connection requirements, refer to the SideCom Communication
System Design and Application Manual (DS059). Contact the Nurse Call manufacturer to make sure the
Nurse Call system is compatibility with the bed.
Classification and Standards
The Progressa+ Bed is designed and manufactured according to the following equipment
classifications and standards:
Flammability Codes—United States, Canada, and Europe
All recommended support surfaces meet the applicable United States, Canadian, and European
flammability specifications.
Siderail Headboard
Footboard Caregiver Pendant
Sleep deck Sleep surface
Line manager Proning accessory
Condition Range
Technical and Quality Standards EN 45501
Classification per EN 45501 Class III
Technical and Quality Assurance Standards ANSI-AAMI ES 60601-1
EN & IEC 60601-1
EN & IEC 60601-1-6
EN & IEC 60601-2-52
EN & IEC 62304
EN & IEC 62366
CAN/CSA-C22.2 No. 60601-1
CAN/CSA-C22.2 No. 60601-2-52
RoHS Directive 2011/65/EU as amended by (EU)
2015/863
Equipment Classification per EN 60601-1 Class I equipment, internally powered equipment
Degree of Protection Against Electric Shock Type B
Classification According to Directive 93/42/EEC Class Im
Class IIa for Advance Surface and Accelerate
Surface (air surfaces)
Degree of Protection Against Ingress of Water Ordinary Equipment - IPX4
Degree of Protection Against the Presence of
Flammable Anesthetic Mixtures
Not for use with flammable anesthetics.
Mode of Operation (Bed Articulation) Continuous operation with intermittent loading,
2 minutes ON/18 minutes OFF
Sound Level <65dBA
Application Environments Environments-1, 2, 3, and 5 per EN and
IEC 60601-2-52