Progressa+ Bed Instructions for Use (216321 REV 2) 65
Rotation Reminders:
Rotation therapy will be suspended when:
Any siderail is lowered. To restart rotation - raise siderail to the up and locked position.
Head of Bed (HOB) is raised higher than 40 degrees. To restart rotation - lower HOB.
Foot of Bed (FOB) is lowered more than 30 degrees. To restart rotation - raise FOB.
Chair position is attempted. To restart rotation - exit chair position.
Percussion/Vibration, P-Max inflate, or Turn Assist is active.
A message will show on the GCI screen when therapy has been suspended for any of the above
conditions.
If CPR is activated, rotation therapy automatically stops and P-Max Inflate is activated. If P-Max
Inflate is active for 60 minutes, the surface will return to normal mode and not the previous therapy
mode.
Check the GCI screen if you are uncertain why the bed is beeping - the reason will be displayed on
the GCI screen.
Setup
Start Rotation
1. Put the patient on the bed.
2. Align the shoulders with the shoulder position label located on the inside of
the head-end siderail.
NOTES:
If Rotation Therapy is desired with Bed Exit on, Bed Exit must be turned on before Rotation
Therapy is started. Only Out of Bed Mode will work during Rotation Therapy.
Bed Exit Position Mode (most sensitive) or Exiting Mode (middle mode) is active, Bed Exit will
switch to Out of Bed Mode (least sensitive) during Rotation Therapy. Once the therapy is
complete the bed will re-arm to the initial Bed Exit Alert Mode.
1. Press the Pulmonary Therapy menu control.
2. Press Rotation.