156 BD FACSCanto II Instructions for Use
Emptying the Waste Container
Figure 10-1 Location of waste port on fluidics cart
1 Ensure the cytometer is not acquiring events.
2 Detach the waste container’s sensor line and fluid line from the fluidics cart
waste port (Figure 10-2 on page 157).
• Pull the sensor straight out.
• Press the metal clip on the fluid line.
All biological specimens and materials coming into contact with them can
transmit potentially fatal disease. To prevent exposure to biohazardous
agents, expose waste container contents to bleach (10% of total volume)
before disposal. Dispose of waste in accordance with local regulations. Use
proper precautions and wear suitable protective clothing, eyewear, and
gloves.
disposable
waste cap
trap
waste
container
waste port