158 BD FACSCanto II Instructions for Use
5 Add approximately 1 L of bleach to the empty waste container (10 L
container).
6 If one month has passed since the last cap change:
• Detach the cap from the trap.
• Replace it with a new cap.
• Write the date on the cap label.
7 Screw the cap assembly onto the waste container and hand-tighten until it
is fully closed.
8 Re-attach the sensor line and fluid line.
Purging the Fluidics Filters
NOTICE The cytometer must be on (system pressurized) during this procedure.
1 Run the Prime After Tank Refill for all fluids to ensure all fluid lines are
primed.
See Priming Fluidics Lines on page 169.
2 Place a few paper towels beneath the filter to be purged.
To prevent waste container overpressurization, do not overtighten the trap
or attached filter cap. Tighten each component only until it is hand-tight.
Do not use sealants such as Teflon® tape or other adhesives.
338677 Rev A
Waste (A)
space for date