Chapter 11: Troubleshooting 225
Display Troubleshooting
Observation Possible Causes Recommended Solutions
Software screen appears
distorted (height is
compressed; New Lot ID
button is missing from
Cytometer Setup
Wizard)
• Display setting was
changed from 96 dpi to
120 dpi
1 Right-click the desktop and
select Properties from the
menu.
2 Click the Settings tab.
3 Click Advanced.
4 Click the General tab.
5 Select Normal Size (96 DPI)
from the DPI Setting menu.
6 Click OK to restart the
computer.