Chapter 11: Troubleshooting 219
BD FACSDiva Software Cytometer Setup
Observation Possible Causes Recommended Solutions
Error creating
compensation tubes
Naming conflict with
existing tube or
worksheet
1 Locate the tube or worksheet
named (ParameterName)
Stained Control, and change
the name.
2 Create the compensation
tubes again.
Error calculating
compensation
PMT voltages
inconsistent between
compensation tubes
Re-record all compensation
tubes with the same PMT
settings.
Wrong tube run for
stained control tube
Re-record all compensation
tubes and recalculate.
Insufficient separation
between positive and
negative populations
Adjust compensation to see if it
improves. If it does not, re-
record the compensation tubes,
draw new gates, and calculate
compensation again.
If automatic compensation fails,
perform manually.