Reusing the Analysis
Now that the analysis strategy has been defined, use it to analyze the remaining
tubes in the experiment. Global worksheets allow you to apply an analysis
strategy to a series of data files without saving the analysis each time. After
previewing the data, you can print the analysis or save it to a tube-specific
(normal) worksheet (see the following section, Saving the Analysis).
1 Move the current tube pointer to the next tube under the LWB specimen.
2 View the data on the global worksheet; make adjustments to gates, as
needed.
Adjustments will also apply to the next tube viewed on the global
worksheet. If you don’t want to alter the global worksheet, save the
analysis as described in the next section and make adjustments on the
tube’s normal worksheet.
Saving the Analysis
Since the analysis objects were created on a global worksheet, the analysis will
not be saved with each tube. If you want to save a copy of the analysis with any
tube, do the following.
1 Expand the TBNK Analysis global worksheet in the Browser.
2 On the Worksheet toolbar, click the Global Worksheets button ( ) to
switch to the normal worksheet view.
3 Create a new normal worksheet for the destination tube; rename the
worksheet.